The list should be of class opiKineticStimulus and contain the
following elements.
path
list of (x,y) coordinates in degrees that is usable by
xy.coords()
image
image[i]
is the image to display (in a machine specific
format) in the section of the path specified by path[i]..path[i+1]
.
levels
if is.na(image)
then levels[i]
is the stimulus
level in cd/\(\mbox{m}^2\) in the section of the path specified by
path[i]..path[i+1]
sizes
sizes[i]
is the size of stimulus (diameter in degrees)
to use for the section of path specified by path[i]..path[i+1]
, or a
scaling factor for images[i]
.
colors
colors[i]
is the color to use for the stimulus in the
section of path specified by path[i]..path[i+1]
. Ignored if !is.na(image).
speeds
speeds[i]
is the speed (degrees per second) for the
stimulus to traverse the path specified by path[i]..path[i+1]
.
...
machine specific parameters
Octopus 900
x
and y
are in degrees, with precision to three decimal places
recognised.
image
is not possible on an Octopus 900.
levels
are in cd/\(\mbox{m}^2\), and are rounded to the nearest
one tenth of a dB for display.
colors
are ignored. Use opiSetBackground()
to alter stimulus color.
sizes
are in degrees, but are rounded to the nearest Goldmann Size I..V
for display.
Kowa AP 7000
Only a simple path with a start and an end point is supported by the AP-7000.
x
and y
are in degrees and should only be length 2. (precision?)
image
is not possible on an Kowa AP 7000.
levels
are in cd/\(\mbox{m}^2\) in the range 0.03 to 3183, and are
rounded to the nearest one tenth of a dB for display. (precision?)
colors
one of .OpiEnv$KowaAP7000$COLOR_WHITE
,
.OpiEnv$KowaAP7000$COLOR_GREEN
, .OpiEnv$KowaAP7000$COLOR_BLUE
, and
.OpiEnv$KowaAP7000$COLOR_RED
.
sizes
are in degrees, but are rounded to the nearest Goldmann Size I..V
for display.
speeds
are in degrees per second in the range 3 to 5.
Compass