Reads the instrument traces, the keys and values describing each module, and
the tables, such as a pump's timetable, from 'Agilent ChemStation' register
(.REG) files. Users reach it through read_agilent_d with
what = "instrument" and through read_chemstation_method; this page
documents the file format.
read_chemstation_reg(path)A list of:
traces: one row per point of each instrument trace, with columns
trace (its title, e.g. "PMP1, Pressure"), unit, time (in minutes)
and value.
conditions: the keys and values stored in every object of the file, with
columns object (its title), key and value (as character). Each
module's Start/Stop Conditions object (e.g.
"PMP1, Start/Stop Conditions") records its state at the start and end of
the run, such as StartPressure and StopPressure for the pump or
ActInjVolume for the autosampler. Every object records the start of the
run as DateTime, and the object of a solvent trace gives the solvent's
name, where one was entered, as Description.
tables: a list of data frames, one per table, named by the key holding
it, such as TIMETABLE for a pump's timetable.
Path to a .REG file.
All integers are little-endian and offsets are counted from 0.
Header. Bytes 02 33 32 00, then the length-prefixed string
REGISTER FILE at 0x04 and a length-prefixed container version at 0x18:
notused in files from revisions B and C, A.xx.xx in files from
revision A. At 0x22 is a u32 offset of the object index and at 0x26 a u16
object count. The index holds one 8-byte entry per object, starting with the
u32 offset of the object; the index itself begins where the last object
ends.
Revision B and C objects. A 12-byte preamble, whose contents differ by
revision, then a stream written by the 'MFC' CArchive class. Each object
starts with a u16 tag: FFFF introduces a new class (u16 schema, u16 name
length, ASCII name); 8000 plus an index refers to a class already read;
7FFF is followed by a u32 tag for large indices. Class indices restart in
every object. The classes are:
CHPUserObject and CHPLCObject: two objects (the second a list of
keys and values), then a flag byte and, if it is set, a data object.
CHPList: one object. CObArray: a u16 count (FFFF then a u32 count
when large) followed by that many objects.
CHPNdrDouble, CHPNdrString and CHPNdrObject: a key, 6 bytes, then a
float64, a string or a flag byte and object. Strings are a u16 length
followed by UTF-16LE; a length of C000 is instead followed by a u16 id
naming a standard key (0 ObjClass, 1 Title), and 8000 by a u16
index, counted from 0, into the non-empty keys and table column names
already read in the object. String values are not counted.
CHPDatLongSliced and CHPDatDoubleSliced: 6 bytes, then the y row and
the x row.
CHPDatLongRow and CHPDatDoubleRow: 14 bytes, the unit string, u16 7,
u16, u32 point count n and a u8 flag marking an implicit row. Unless
implicit, n int32 (Long) or float64 (Double) values follow. Then u32,
float64 first value, u8, u32 flag marking scaled values, float64 scale and
float64 offset. Values are raw x scale + offset if scaled and raw
otherwise; an implicit row is first + i x scale.
CHPTable: a header holding a u16 row size at byte 2, a u16 row count n
at byte 4, and the u16 counts of data and extra columns at bytes 20 and 22
(16 and 18 for schema 11). Then an 18-byte descriptor per column (u16
offset in the row, u16 size, u16 type and u16 flags, then 10 bytes);
(n + 1) x row size - 4 bytes holding a default row, less its first 4
bytes, then the n rows, followed by the value of each extra column with
flag 40; and each column's name, a string whose length
counts a terminator. A data column of type 3 with flag 40 is followed by
its default and n strings. Type 4 is float32, with -10000 for an empty
cell, and type 5 int32. The extra columns are not returned.
CHPAnnText is read past but not returned.
In LCDIAG.REG each trace is an object whose Title names it. Revision
B.01 stores traces as CHPLCObject with CHPDatLong rows and an implicit
time axis; B.04 does the same but writes a space before the comma in trace
names ("PMP1 , Pressure"), which is removed; revision C stores them as
CHPUserObject with unscaled CHPDatDouble rows and explicit, possibly
uneven, times. The other objects, titled Start/Stop Conditions, hold only
keys and values.
Revision A objects. One byte, a u32 record count n, n 16-byte record
headers (u16, u16 type, u32 size, u32, u32 id), 4n further bytes, then the
data of each record in turn. Records refer to one another by id. Types
8001 and 8003 are Latin-1 strings and 8006 a string after 2 bytes;
8002 holds a numeric array; 0602 (43 bytes) a key at byte 14 and float64
at byte 35; 0601 a key and the u32 id of its value at byte 35; 0603 the
key of a table and the u32 id of its 0701 record at byte 35. A 0701
record holds a u16 row size at byte 2, a u16 row count n at byte 4, the u32
offset of its rows at byte 6 and a u16 count of data columns at byte 16,
then a 30-byte descriptor per column: a 16-byte name, then the u16 offset
in the row, size, type and flags, as in revision B and C. The rows are
(n + 1) x row size bytes, the first a default row. Types
0501 and 0503 (161 bytes) describe a trace: u32 point count at byte 9,
then the u32 ids of the x unit and data and the x scale at bytes 27, 31 and
61, and those of y at bytes 94, 98 and 128. An array id with no record
denotes an implicit axis of i x scale.
Not known: the meaning of the skipped bytes, and whether the scale or the offset comes first in a row (every file seen has an offset of 0).
Ethan Bass
Objects that cannot be parsed are skipped with a warning. Tested on
LCDIAG.REG files from revisions A.10.02, B.01.03, B.04.02, B.04.03,
C.01.03, C.01.07 and C.01.10, and on method register files from four
revision B and C instruments and two revision A instruments.