Uncertainty in a given balance reading considering the effects of
rounding error, lack of repeatability, eccentricity and balance taring.
Usage
uncertReading(calibCert, reading, units = NULL, sd = NULL,
sd.units = NULL, d = NULL, d.units = NULL)
Arguments
calibCert
object of class "calibCert" with the calibration
data of the balance. See calibCert() for details.
reading
numeric with balance reading for the mass of the object.
units
character with the units of reading. Must be a SI unit.
If not provided, the balance standard units are assumed.
See calibCert() for details.
sd
standard deviation when a balance reading is the result
of averaging several individual measurements of same object. If not provided
the information is taken from the calibration certificate of the balance
sd.units
character with the units of standard deviation.
If not provided, the value stated at units or the balance
standard units is used.
d
balance division scale. Useful when the balance is operated a a
division scale different from that stated in the calibration certificate.
This is the common case when the user is give up some readability
in order to make faster mass measurements. If not provided.
the functions uses
the balance division scale stated in the calibration certificate.
d.units
character with the units of parameter d.
If not provided, the value stated at units or the balance
standard units is used.
Value
A numeric value of uncertainty for a balance reading.